DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Catalog Number |
MA-D20 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and the reported condition that the axis of the attachment device could not move was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was determined that the device had heat.The assignable root cause of this condition was determined to be traced to maintenance, which is improper maintenance.
|
| |
|
Event or Problem Description
|
|
It was reported that the axis of the minimal access attachment device could not move.During in-house engineering evaluation it was determined that the device had heat, a worn bearing, foreign substance/debris/cleaning/sterilization, and corrosion/rusting/pitting.It was further determined that the device failed pretest for visual assessment and temperature assessment.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|