• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number MA-D20
Medical Device Problem Codes Mechanical Jam (2983); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and the reported condition that the axis of the attachment device could not move was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was determined that the device had heat.The assignable root cause of this condition was determined to be traced to maintenance, which is improper maintenance.
 
Event or Problem Description
It was reported that the axis of the minimal access attachment device could not move.During in-house engineering evaluation it was determined that the device had heat, a worn bearing, foreign substance/debris/cleaning/sterilization, and corrosion/rusting/pitting.It was further determined that the device failed pretest for visual assessment and temperature assessment.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINIMAL ACCESS ATTACHMENT
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key22100171
Report Number1045834-2025-00352
Device Sequence Number7343268
Product Code HBE
UDI-Device Identifier00845384004356
UDI-Public(01)00845384004356
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K042783
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMA-D20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Initial Report FDA Received Date05/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-