Catalog Number SXPP1B406
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Wound Dehiscence (1154)
Date of Event 05/08/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the sales rep that a patient underwent a robotic hysterectomy in (b)(6) 2025 and barbed suture was used.Four months post op the patient had sexual intercourse and felt a pop.The patient is a breast cancer survivor.Patient has been having a lot of sex since being cleared.Was having a lot of pain and bleeding.Patient was seen in the office and noted to have a vaginal cuff dehiscence.Repair done on (b)(6) 2025.Used different sutures to repair.None of the barbed suture was left at all.Additional information has been requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: confirmed the pop experienced by the patient 4 months post-op was the suture breaking.The sales rep reported that stratafix spiral suture sxpp1b406 may also been used on the patient.The surgeon always does 2 back stitches with stratafix.Post op instructions for all patients includes no intercourse for 8 weeks and examination prior to clearance.No estrogen prior to healing.Lot number is unknown.No additional information is available.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.What is meant by ¿leaving a hole behind.¿? did the barbs not hold in the tissue post op? did the stratafix suture break? if so, where was the break noted (termination, middle, end)? did the suture pull out of the tissue? were there any patient stress factors that led to the suture breaking or pulling out of the tissue? if not already given, please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.Date and name of index surgical procedure? patient 1:(b)(6) 2025 - robotic hysterectomy with bilateral salpingectomy, right oophorectomy, extensive excision of endometriosis, left ureterolysis patient 2: (b)(6) 2025 robotic hysterectomy patient 3: unknown date ¿ unknown procedure the diagnosis and indication for the index surgical procedure? what was the initial approach for the index surgical procedure? (open, laparoscopic or other)? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? was the fixation loop secured to tissue at the initiation of suture use during the index procedure? was at least one reverse stitch performed prior to closure? please describe the appearance of the suture during the second procedure.Onset date/time of dehiscence? (# post op days) please describe any surgical intervention required for the wound dehiscence including date and findings.Did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? what symptoms did the patient experience following the index surgical procedure? onset date? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? lot number?.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFXSPI PDS+ UNI VIO 6IN 0 SA CT-1
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
*   00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22100646
Report Number2210968-2025-05957
Device Sequence Number15289497
Product Code NEW
UDI-Device Identifier10705031236097
UDI-Public10705031236097
Combination Product (Y/N)Y
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K150670
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1B406
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2025
Initial Report FDA Received Date05/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexFemale
Patient Weight136 KG
-
-