• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON TORIC ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON TORIC ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CCA0T3
Medical Device Problem Code Device Difficult to Setup or Prepare (1487)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2025
Type of Reportable Event Malfunction
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, it was noted that the lens haptic malfunctioned/ did not reach patient.There was no patient contact.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The product was returned for analysis and the reported complaint could not be observed.Iol was not returned.Only the device was returned.The device was returned in an opened blister tray inside the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle.No damage observed.The product investigation could not identify the root cause for the reported complaint "haptic malfunctioned/ did not reach patient " as the lens was not returned.Only the device was returned for evaluation.Due to this, we are unable to confirm the lens and the plunger position for advancement and determine a root cause.The reported complaint is lacking in relevant information such as viscoelastic used to complete a thorough investigation.Due to the lack of information, we are unable to verify if the product contributed to the event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22105280
Report Number9612169-2025-01147
Device Sequence Number15234449
Product Code MJP
UDI-Device Identifier00380652440158
UDI-Public(01)00380652440158(17)280411(11)240413(10)25833909
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCA0T3
Device Lot Number25833909
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/07/2025
Supplement Date Received by Manufacturer07/15/2025
Initial Report FDA Received Date05/29/2025
Supplement Report FDA Received Date08/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/13/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-