| Catalog Number |
802516 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was raised bumps on the shaft of the intermittent catheter.Patient experienced pain.Lot#ngjy3625, lot#ngjx1963, lot#ngjx3142.No medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive as no sample was returned for evaluation.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.A potential root cause for this failure mode could be rough surface, insertion difficulty due to rough drainage eye.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: indications for use: for urological use only.Single use.Single patient use.Do not reuse.Do not use if package is open or damaged.Sterilized using ethylene oxide storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Warnings: on latex and red rubber catheters, do not use ointments or lubricants having a petroleum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristic of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local state and federal laws and regulation.Adverse reactions: urinary tract infection; bleeding from urethra; irritation of the urethra.Instruction for intermittent catheters: wash your hands thoroughly with soap and water.Remove catheter from the pack.Position yourself comfortably cleaning the opening of the urethra and surrounding area.Gently insert rounded end of catheter into urethra.When urine stops flowing, remove catheter from urethra.Dispose of catheter in accordance with local rules and regulations.Wash your hands.Correction: d.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that there was raised bumps on the shaft of the intermittent catheter.Patient experienced pain.Lot#ngjy3625, lot#ngjx1963, lot#ngjx3142.No medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|