• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. EPSILA EV¿; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. EPSILA EV¿; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number EAZ101
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Perforation (2001)
Date of Event 05/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: product id ev240152 lead implanted: (b)(6) 2025, explanted: (b)(6) 2025.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient complained of feeling blunt pain all around and under the sternum three hours post implant of an ex travascular (ev) lead.It was noted that the patient experienced muscle stimulation and felt the subthreshold lead impedance checks.Pleural placement of the lead was suspected due to slight lateral movement of the distal part of the lead with breathing on fluoroscopy.High pacing threshold was also observed.A computerized tomography (ct) scan was performed two days later which showed the distal part of the lead was placed inside the pleural space.During the revision, noise and pause prevention episodes were noted.The extravascular (ev) lead was extracted and replaced.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPSILA EV¿
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22106555
Report Number2182208-2025-02618
Device Sequence Number7122746
Product Code NVY
UDI-Device Identifier00763000369736
UDI-Public00763000369736
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P220012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEAZ101
Device Catalogue NumberEAZ101
Device Lot Number102406525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Initial Report FDA Received Date05/29/2025
Date Device Manufactured03/06/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H11...."
Outcome Attributed to Adverse Event Hospitalization; Life Threatening;
Patient Age37 YR
Patient SexFemale
-
-