Catalog Number 7400-131-000
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer field service technician reported that the device had missing blade retainer while it was being serviced at the user facility.There was no patient involvement, no delay, no medical intervention, and no adverse consequences.
 
Additional Manufacturer Narrative
Additional information: d9.Device evaluation: follow-up report submitted to document device evaluation results.Correction: upon further investigation, it was determined that the blade retainer is a large component and is unlikely to result in a death or serious injury.This event is not reportable.
 
Event or Problem Description
The manufacturer field service technician reported that the device had missing blade retainer while it was being serviced at the user facility.There was no patient involvement, no delay, no medical intervention, and no adverse consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPX XL OSCILLATING SAW
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
kelsey barla
1941 stryker way
portage, MI 49002
2693237700
MDR Report Key22106711
Report Number3015967359-2025-01058
Device Sequence Number8140089
Product Code ERL
UDI-Device Identifier07613327483024
UDI-Public(01)07613327483024
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K112593
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number7400-131-000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2025
Supplement Date Received by Manufacturer07/21/2025
Initial Report FDA Received Date05/29/2025
Supplement Report FDA Received Date07/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-