The device was not returned thus a proper device analysis could not be completed nor the reported issue confirmed.This lens was implanted using a non-validated injection system as well as utilized an off-label procedure (yamane technique).There was no damage reported to have been noticed during the inspection prior to use and preparation, which suggests a product deficiency, did not contribute to the reported difficulties.Based on the information provided, the risk assessment for the lucia product, and our experience we have determined that the off-label use of the yamane technique caused this incident.The device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lens that would have caused or contributed to the nature of this complaint.Additionally, our lenses are 100% inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operating procedures and inspections and have met all criteria for release.
|