| Catalog Number |
NSLG2C35A |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/29/2025.B3: event year only reported: 2025.D4 batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure jaws not opening mid case.There were no patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/10/2025.D4 batch #: a97d8v.Corrected data: h4.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with tissue stuck in the jaw.Excessive body fluids may influence the opening and closing of the jaws.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All three tones were heard during functional testing (tone 1 is heard when the energy activation button is pressed; tone 2 is heard when the tissue impedance threshold is reached and tone 3 is heard when the cycle is complete).No issues were noted with opening and closing of the jaws.The accumulation of body fluids is a routine occurrence during surgical procedures and does not necessarily indicate a manufacturing defect in the device.There were no anomalies noted with the functionality of the device.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch a97d8v, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|