| Catalog Number |
EP-200148 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Joint Dislocation (2374)
|
| Date of Event |
04/08/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately two months post-implantation, the patient underwent a hip revision due to infection and dislocation.The patient had a washout and the bearing and head were exchanged.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat: 00877502802, lot: 3175454 bioloxâ® delta, ceramic femoral head.G2: foreign ¿ australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation, as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: d4; d5; g3; h2; h3; h4; h6 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined for the dislocation.Single-use, sterilized devices manufactured or distributed by zimmer biomet are sterilized in accordance with fda regulations and iso standards to a sterility assurance level of 1.0 x 10-6 or better.Therefore, it is unlikely that the specified device caused any patient infection.This complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|