A signal loss was reported with the abbott diabetes care (adc) device in use with iphone ios operating system 18.4.1 and the customer was unable to obtain glucose readings and receive glucose alarms.As a result, the customer was not alerted of changes in glucose level and presented at a medical facility.Upon arrival, an unspecified reading was obtained on a healthcare professional (hcp) meter and the customer received hcp administration of an unspecified infusion as treatment for the diagnosis of hyperglycemia.There was no report of death or permanent impairment associated with this event.
|
|
An investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The customer experienced signal loss with the reported application.After attempting to identify the customer's reported configurations, the information provided in the complaint was insufficient to conduct a comprehensive investigation.The lack of the iphone model restricts the ability to identify potential causes of the customer's reported issue.Therefore, the reported issue is not confirmed due to the inadequate information provided.The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The device mfg date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|