| Model Number |
BROACH, PEGGED GLENOID |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 05/12/2025, it was reported by a sales representative via (b)(4) that an ar-9233 broachs end broke off where it connects to the impactor pad.This occurred during use in a case with no patient effect.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.Upon visual inspection, it was noted that the pin.125 was broken from the handle, and the anvil was broken off.Functional testing was not performed due to the damage to the device.The most likely cause of this failure is attributed to wear and tear damage incurred over repeated usage in the field.Manufacturing date 2015.
|
| |
|