MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752439
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that the aspiration failure of the probe occurred, and the vitrectomy cutter could not aspirate during vitrectomy surgery.The surgery was completed on the same day after replacing the vitreous cutter with another one.There was no patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The probe was visually inspected.The returned probe aspiration line at the entrance of the tubing insertion to the probe engine inside the rear shell was found kinked.The folded aspiration line prevented the sample from priming on the console and causing aspiration failure.The light emitting diode (led) rings on the console turned green as the probe connectors were engaged to the console indicating the proper communication between the probe and the calibrated console, 10000 cut rate displayed on the console screen.The kink on the tubing happened during assembling the rear shell to the probe engine.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.¿ the most likely root cause for the customer¿s event is a manufacturing defect.The vitrectomy probe has a rear shell meant to improve the ergonomics of the handpiece, however, if the rear shell is improperly installed, it will result in the customers reported event.¿ action will not be taken based on this occurrence.An analysis of similar complaints confirmed no unfavorable trend for this event.Routine training is performed to ensure non-conforming product is rejected and placed in a reject bin during production.Quality assurance will continue to monitor and will take action for future occurrences as is deemed necessary.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22112122
Report Number1644019-2025-01633
Device Sequence Number7063093
Product Code LRO
UDI-Device Identifier00380657524396
UDI-Public(01)00380657524396(17)260228(11)240312(10)16L8HA
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue Number8065752439
Device Lot Number16L8HA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer08/11/2025
Initial Report FDA Received Date05/30/2025
Supplement Report FDA Received Date09/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONSTELLATION VISION SYSTEM
Patient SexUnknown
-
-