• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL SA UNIVERSAL WIREDRIVER MIC-MTRC 1-4MM; SURGICAL POWER TOOL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL SA UNIVERSAL WIREDRIVER MIC-MTRC 1-4MM; SURGICAL POWER TOOL Back to Search Results
Catalog Number 89-8509-451-41
Medical Device Problem Codes Difficult to Insert (1316); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the wire driver pin cannot be connected, inserted, or held in place.No consequences or impact to the patient.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).G2 ¿ foreign ¿ india.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: b4, b5, d4, d9, e1, g3, g6, h1, h2, h3, h6, h11.The following section was corrected: h6 - component codes.No product was returned or pictures provided; visual and functional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Review of complaint history identified additional similar complaints for the reported item and no additional similar complaints for the reported part and serial combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL WIREDRIVER MIC-MTRC 1-4MM
Common Device Name
SURGICAL POWER TOOL
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ   CH-1228
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key22112840
Report Number0008031000-2025-00069
Device Sequence Number7063119
Product Code HAB
UDI-Device Identifier00889024354951
UDI-Public(01)00889024354951(11)230427(10)5018561
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number89-8509-451-41
Device Lot Number5018561
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2025
Supplement Date Received by Manufacturer06/19/2025
Initial Report FDA Received Date05/30/2025
Supplement Report FDA Received Date07/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-