MAUDE Adverse Event Report: BIOMET 3I 3I T3 SHORT EXTERNAL HEX PARALLEL WALLED IMPLANT WITH DCD, 5MM(D) X 5MM(L); DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I 3I T3 SHORT EXTERNAL HEX PARALLEL WALLED IMPLANT WITH DCD, 5MM(D) X 5MM(L); DENTAL IMPLANT Back to Search Results
Catalog Number BNES505
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Sinus Perforation (2277)
Date of Event 01/09/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.
 
Event or Problem Description
The doctor reports that the dental implant located in position number #46 perforated the sinus during implantation and was removed to avoid sinus penetration.The doctor reports that the procedure was not concluded by placing another implant.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4) the following fields have been updated: b4: date of this report.B5: describe event or problem.D4: additional device information.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) bnes505, (3i t3® short external hex parallel walled implant with dcd®, 5mm(d) x 5mm(l)) for evaluation.Visual evaluation of the as returned device identified the implant with signs of use and was attached to its bundle mount.However, no apparent malfunction that would cause or contribute to the event was identified.Measurements match drawing.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number 2021090929.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2021090929 for similar events and no other complaint was identified.Review completed utilizing keywords: perforated sinus.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are patient factors and improper techniques used.Therefore, based on the available information, a device malfunction did not occur.The reported event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 SHORT EXTERNAL HEX PARALLEL WALLED IMPLANT WITH DCD, 5MM(D) X 5MM(L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22114236
Report Number0001038806-2025-01235
Device Sequence Number17206294
Product Code DZE
UDI-Device Identifier00844868033592
UDI-Public(01)00844868033592(17)261007(10)2021090929
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K150571
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue NumberBNES505
Device Lot Number2021090929
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/30/2025
Supplement Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age73 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-