ARTHREX, INC. FIBERTAK; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number FIBERTAK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 14th may 2025, it was reported by an arthrex employee via email that for an ar-3638, knotless fibertak¿ with #2 suture (5-box), its implant suture broke when the surgeon pulled to reduce the repair suture.This was detected during a shoulder instability repair procedure on (b)(6) 2025.The procedure was completed successfully by using one more suture anchor.There was no patient harm.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening/tensioning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAK
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22115005
Report Number1220246-2025-02294
Device Sequence Number17186393
Product Code MBI
UDI-Device Identifier00888867292123
UDI-Public(01)00888867292123(11)241127(17)291031(10)15354146
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K130458
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFIBERTAK
Device Catalogue NumberAR-3638
Device Lot Number15354146
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2025
Supplement Date Received by Manufacturer10/02/2025
Initial Report FDA Received Date05/30/2025
Supplement Report FDA Received Date10/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-