The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information provided, a definitive cause for the reported difficulty could not be determined.Possible factors that can contribute to difficult to remove/resistance with the introducer sheath post-inflation include, but are not limited to, loose balloon folds, guiding catheter lumen obstructions, kinks, bends, procedural technique, insufficient support or interactions with other devices.In this case, a conclusive cause for the reported difficulty removing the device could not be determined since the device or component were not returned for analysis.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional device referenced in b5 is filed under separate medwatch report number.
|