• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR ACUMEN IQ CUFF ADULT; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR ACUMEN IQ CUFF ADULT; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE Back to Search Results
Model Number AIQCA
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use of the acumen cuff it displayed a map value approximately 20-30 mm hg lower than expected when compared to the nibp.Troubleshooting did not resolve this issue.There was no treatment provided based on inaccurate values and no patient injury.
 
Additional Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming when the investigation is completed.The device history record review has not yet been completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Additional Manufacturer Narrative
One adult acumen iq finger cuff was returned for evaluation.The customer report of inaccurate values was unable to be confirmed.Finger cuff passed eeprom verification with expired status and patient information.Finger cuff was reset for pressure test.Finger cuff zeroed and sensed pressure on hemosphere monitor without any error message.Electrical testing showed that all elements such as shield, led, and photodiode of returned unit were ok.No visible damage was noticed from the returned finger cuff.Finger cuff sensor passed both pre and post-decontamination leak test.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.Since the complaint affected unit was returned for evaluation and no defect was found a product non-conformance or device failure associated to manufacturing or design could not be confirmed.The instructions for use contains the following warnings: improper placement, sizing, or alignment of the finger cuff can lead to inaccurate monitoring.Do not apply finger cuff(s) on a hand or finger when a second blood pressure measurement device is actively monitoring on the same arm (or hand or finger).As part of the manufacturing process 100% of the units go through an electrical testing, visual inspection, and qa sampling inspection.A device history record review was completed and documented that device met all specifications upon distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUMEN IQ CUFF ADULT
Common Device Name
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES
17200 laguna canyon rd
irvine CA 92618
Manufacturer Contact
erika bonilla
1 edwards way
irvine, CA 92614
MDR Report Key22116249
Report Number2015691-2025-04304
Device Sequence Number15247710
Product Code DXN
UDI-Device Identifier00690103219067
UDI-Public(01)00690103219067(17)261120(11)241120(10)66150652
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K230919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAIQCA
Device Lot Number66150652
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2025
Supplement Date Received by Manufacturer08/26/2025
Initial Report FDA Received Date05/30/2025
Supplement Report FDA Received Date08/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-