| Catalog Number |
13546-02 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.E1: the reported phone number is "(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during an ent procedure, the coblator ii controller was not generating plasma and the wand connector was burned.The procedure was successfully completed with a surgical delay greater than 30 minutes using a smith and nephew back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was received for evaluation.The visual inspection was performed on the complaint unit and found that the wand port burnt.The functional inspection was performed on the complaint unit.It was tested with known good wand port receptacle assy and its working properly.Hence it was confirm that wand port receptacle assy was faulty.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with a component failure.Factors that could have contributed to the reported event include a wand detection circuit failure, damaged receptacle, plugging in a wet wand connector, or exposure of saline to the wand receptacle.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|