MAUDE Adverse Event Report: CYBERONICS - HOUSTON PULSE GEN MODEL 105; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON PULSE GEN MODEL 105; GENERATOR Back to Search Results
Model Number 105
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Convulsion, Clonic (2222)
Date of Event 05/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that recently the patient began to experience an increase in seizures.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
 
Additional Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event or Problem Description
Response was received from the physician.The increase in seizures was assessed to be due to low battery and seizure levels were back to pre-vns levels.Implant card was also received reporting a generator replacement due to prophylactic reasons.The suspect product has not been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 105
Common Device Name
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics boulevard
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics boulevard
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
houston, TX 77058
2816672681
MDR Report Key22117605
Report Number1644487-2025-10186
Device Sequence Number4109924
Product Code LYJ
UDI-Device Identifier05425025750054
UDI-Public05425025750054
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 06/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date06/14/2015
Device Model Number105
Device Lot Number202623
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/30/2025
Supplement Report FDA Received Date06/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age17 YR
Patient SexMale
-
-