ARTHREX, INC. QUADLINK IMPLANT SYSTEM, 10MM; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number QUADLINK IMPLANT SYSTEM, 10MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 05/12/2025, a sales representative reported via (b)(4) that an ar-1288qis-100 quadlink implant systems implant broke when the surgeon was preparing the graft.The second complaint is that the quadpro harvester was dull and was not able to cut the quad tendon for harvest.This occurred during an autograft quadriceps acl reconstruction case.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6.
 
Event or Problem Description
Additional information was provided on 05/28/2025, where the sales representative stated the harvester did not break as it was simply dull and would not cut.The fibertag suture broke when the surgeon was preparing the graft and pulling on it.The patient's bone quality was good.No instrument was used to prepare the bone socket.All fragments were retrieved from the patient.They opened another quadlink kit to complete the procedure.A small delay occurred, no more than 2 minutes.No additional anesthesia was administered.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to prying/leveraging the device during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADLINK IMPLANT SYSTEM, 10MM
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22117629
Report Number1220246-2025-02312
Device Sequence Number4084159
Product Code MBI
UDI-Device Identifier00888867344754
UDI-Public(01)00888867344754(11)250331(17)280430(10)15412587
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K241235
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQUADLINK IMPLANT SYSTEM, 10MM
Device Catalogue NumberAR-1288QIS-100
Device Lot Number15412587
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer05/28/2025
09/30/2025
Initial Report FDA Received Date05/30/2025
Supplement Report FDA Received Date06/25/2025
10/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was