|
On (b)(6) 2025, a customer reported to hologic receiving discrepant results using the aptima hpv (human papillomavirus) assay master lot 908328 on panther plus instrument sn# (b)(6).Sample ids # (b)(6) were part of a cluster and initially resulted hpv positive in hpv worklist # (b)(4).Per customer¿s lab sop, if there is a cluster of three hpv positives containing results with an s/co (signal-to-cutoff) value < 5, the samples with s/co<5 will be retested, and the retested result will be reported out.Upon retest, sample id # (b)(6) resulted hpv negative, and this result was reported out.Sample ids # (b)(6) were not retested as these samples resulted hpv positive with an s/co > 5.No patient treatment information was provided by the customer.Hologic has not learned of any adverse patient outcomes due to this event.
|