|
H6: the device was evaluated by ventec where the reported issue of it delivering an erratic (low) vte (exhaled tidal volume) level was confirmed, however, the reported issue of the device being unable to maintain peep (positive end expiratory pressure) could not be confirmed.Ventec did observe that the device's breath rate was mildly higher than expected but was unable to confirm that it was "erratic".Ventec then conducted a review of the devices settings and its system logs which showed sequential ventilation therapy preset changes without performing pre-use tests (put) in between.When the device preset was changed without a put being performed, ventec began to observe unstable device performance.Ventec determined that this behavior was caused by a difference in the humidification setting of the two presets.The first customer preset was using a humidification setting of "hme" whereas the second was using the "humidifier" option.When the device was switched between the two presets, the compensation that occurred for whichever humidification setting was invalidated.A put is mandatory between preset swaps if the customer continues to use the alternative humidification settings.To confirm this setting was the cause of the customer¿s issues, the customer's presets were both set to hme.An initial put was performed, and the device was allowed to ventilate on preset one.The device was then swapped to preset two (now using hme), and it ventilated as expected.There was no unexpected device performance during or after the transition between presets.Proper device operation was confirmed through functional and performance testing.The vocsn clinical and technical manual [pg.50] states the following: "the vocsn pre-use test calculates the resistance and leak of the ventec one-circuit.Based on these calculations, vocsn verifies the integrity of the ventec one-circuit, and also improves the accuracy of therapy delivered during ventilation.If used, the pre-use test will also verify the connection status of the ventec one-circuit o2 tube.To ensure therapy is delivered accurately, you must perform a pre-use test whenever prompted, the patient circuit is changed, or the device is powered on." "some control changes cause vocsn to prompt you to run a pre-use test.You may also press the menu tab and then the pre-use test button to begin a pre-use test at any time." the investigation determined that the cause of the reported issue was user error, as the user was not performing a pre-use test (put) when changing between presets.
|