MDT PUERTO RICO OPERATIONS CO, MED REL CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 55840006535
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 04/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received via healthcare provider via manufacturer representative regarding product used for transforaminal lumbar interbody fusion (tlif) surgery.It was reported that the screws were broken and fragments left inside the patient.There were no further complications or symptoms reported regarding the event.Additional information was received via manufacturer representative that patient had low back pain and revision surgery is planned.Preoperative diagnosis for this procedure is l5-s1 spondylolisthesis due to isthmic spondylolysis.Additional information was received via manufacturer representative that patient underwent revision surgery to explant the proximal portion of the screw, leaving the entire shaft inside the vertebral body and doing well with no complications now.
 
Event or Problem Description
Additional information was received via manufacturer representative that all the devices that were implanted in initial surgery were explanted but only two screws that were on upper level were broken.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: product analysis of part# 55840006535 lot# h5927835 visual examination of the mas bone screw confirmed bone screw complete fracture at approximately ~3 threads from the base of the bone screw head.Optical examination did not identify material defect near the area of fracture origin, which could contribute to crack propagation.Microscopic examination of the fracture surface identified gently curving convex striations through the cross-sectional area of the implant, which is consistent with overload until subsequent mechanical failure of the implant.Radiographic image: ap x-ray l5-s1 fusion, one of the s1 screws appears as fractured.Lateral x-ray shows both s1 screws appears fractured.Flouro lateral l4-s1 screws present, one screw retained in s1.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key22120435
Report Number1030489-2025-02120
Device Sequence Number15243171
Product Code KWP
UDI-Device Identifier00613994971739
UDI-Public(01)00613994971739(11)240130(10)H5927835
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55840006535
Device Catalogue Number55840006535
Device Lot NumberH5927835
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2025
Supplement Date Received by Manufacturer07/11/2025
08/11/2025
Initial Report FDA Received Date05/31/2025
Supplement Report FDA Received Date08/06/2025
08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexMale
-
-