SMITH & NEPHEW, INC. TWINFIX ULTRA PK 5.5MM W/2 UB-WHT & BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72202599 |
| Medical Device Problem Codes |
Break (1069); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during surgery, the head of a twinfix ultra anchor was disconnected, while implanting it.Surgery was completed with a back-up device, instead.It is unknown if there was any surgical delay, but no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11 h3, h6 the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The insertion device was returned with the anchor and suture strings in place.The anchor is fractured just below the proximal end of the suture window.One of the prongs at the distal end of the insertion shaft is broken off the other is deformed.There is biological debris on the returned items.An evaluation of the customer provided image showed the original packaging with the correct identification information.The insertion device is seen with a fractured anchor attached.The distal tip was separated from the rest of the anchor body.Distal threads look clean cut.Based on the condition of the product material found during visual inspection, it was determined the device damage was caused by the application of improper/excessive force.Additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found tensile strength requirements are specified.A material certificate of analysis (coa) is required for raw material.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event).Based on this investigation, there is no evidence to suggest a design, material or manufacturing issue.Therefore, the need for corrective action is not indicated.
|
| |
|