MAUDE Adverse Event Report: MICROVENTION INC. HEADWAY; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION INC. HEADWAY; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number MC162156S-AL
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation found that an in-house guidewire was able to advance past the hub but became stuck approximately 25cm into the lumen of the returned headway microcatheter.Dried blood from the procedure was found obstructing the lumen.Further investigation found the lumen to not be fully flared at the hub joint, an exposed braid in the lumen, and delaminated ptfe in the lumen, which could have caused or contributed to the guidewire advancement issue.The physical evaluation of the device could not identify the exact conditions or circumstances that led to the lumen flaring issue, exposed braid, and delaminated ptfe, but these conditions are consistent with a deviation in the manufacturing process and will be addressed per the quality system process.A capa has been initiated.A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.
 
Event or Problem Description
It was reported that a headway duo microcatheter would not accept a guidewire.A new catheter was selected and the guidewire worked with it.The device had no patient contact.When the device was received for evaluation, the investigation findings found the lumen to not be fully flared at the hub joint with exposed braid in the lumen, and delaminated ptfe was in the lumen.Two other catheters from the same case experienced the same issue during prep and were reported under mdrs 2032493-2025-90167 and 2032493-2025-90171.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADWAY
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MICROVENTION INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key22121363
Report Number2032493-2025-90212
Device Sequence Number17186445
Product Code DQO
UDI-Device Identifier00842429117408
UDI-Public00842429117408
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K120917
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC162156S-AL
Device Lot Number0000808727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2025
Initial Report FDA Received Date06/01/2025
Date Device Manufactured12/06/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-