MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3341102
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Crushing Injury (1797); Deformity/ Disfigurement (2360)
Date of Event 04/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.H6.Health effect - clinical code e2402: chest injury.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a breast augmentation primary with mentor memoryshape breast implants and experienced rupture on the right side post-operatively, which was confirmed via mri.The patient was involved in a car accident which may have caused or contributed to the event.As a result, the patient underwent removal on (b)(6) 2025.
 
Additional Manufacturer Narrative
On jun 18, 2025, the mentor failure analysis lab received the device for evaluation.On jun 18, 2025, it was noticed the previous report contained an error, stating the right breast implant was ruptured.The left breast implant was the reported ruptured device.On jun 20, 2025, additional information was received indicating the patient also experienced bilateral ptosis.This report is for the left breast prosthesis.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned device revealed that the mentor memoryshape¿ mh 300cc breast implant was found to be ruptured.The evaluation determined that the possible cause of the rupture is the trauma experienced.Rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of gel-filled mammary prosthesis.Some breast ptosis is a normal component of the mature breast.Ptosis becomes undesirable when the breast parenchyma predominates below the areola, droops considerably below the inframammary fold and the nipple points downward.Augmentation alone, without consideration of the ptosis can produce a less than desirable cosmetic result known as a "rock in sock" deformity or an increased ptotic appearance.Ptosis is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22123315
Report Number1645337-2025-06033
Device Sequence Number4251322
Product Code FTR
UDI-Device Identifier00081317005186
UDI-Public(01)00081317005186(17)211130(10)7400044
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Catalogue Number3341102
Device Lot Number7400044
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/13/2025
Supplement Date Received by Manufacturer06/18/2025
Initial Report FDA Received Date06/02/2025
Supplement Report FDA Received Date07/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-