• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number SEP6000
Medical Device Problem Codes Failure to Deliver Energy (1211); Self-Activation or Keying (1557); Failure to Cut (2587); Failure to Fire (2610); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Superficial (First Degree) Burn (2685)
Date of Event 03/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that four devices experienced self-activation without the activation button or switch being pushed during procedure, specifically the cut mode (yellow button) activated itself and resulted in a smaller than 1 cm first degree burn on the patient's arm while coag mode did not activate at all.The burn was cold wrapped for a few minutes.The patient sustained a temporary injury from the burn but recovered well.Other pencils worked fine with the same generator.
 
Additional Manufacturer Narrative
D10 concomitant products: sep6000, pencil sep6000 rkr sw telescoping (lot #: 2405149x), sep6000, pencil sep6000 rkr sw telescoping (lot #: 2405149x), sep6000, pencil sep6000 rkr sw telescoping (lot #: 2405149x).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22123883
Report Number2182208-2025-02651
Device Sequence Number12581280
Product Code GEI
UDI-Device Identifier10884521750913
UDI-Public10884521750913
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K182772
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSEP6000
Device Catalogue NumberSEP6000
Device Lot Number2405149X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2025
Initial Report FDA Received Date06/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-