| Model Number |
CCA0T0 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during intraocular lens (iol) implant procedure, the trailing haptic was stuck to posterior surface of optic.The surgeon tried to manipulate with forceps and also with ia (irrigation and aspiration) nothing worked in bag.The lens was vaulted in anterior chamber and had to peel with forceps.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The product was not returned for analysis.It was reported that haptic was observed to be stuck to the optic after implantation.The root cause of the reported issue is deemed to be related to manufacturing process change that increased the likelihood of company haptics adhering to the posterior optic surface following delivery with the autonome device.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|