ETHICON INC. PROLENE SUTURE 30"(75CM) 6-0 BLU; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number EP8707H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.D4: full udi currently unavailable since the exact lot of the suspected device cannot be determined.
 
Event or Problem Description
It was reported that a patient underwent an anastomosis procedure on (b)(6) 2025 and suture was used.During the procedure, the needle pull-off from suture during sewing the vessel wall, it was hold in the needle holder.No patient harm nor significant delay reported.Procedure finished with same like device.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 7/2/2025 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found additional information: d4, d9, h3, h6 h3 investigational summary: the product was returned to ethicon for evaluation.Two sealed boxes, each one with thirty-six unopened samples, were received for analysis.As per the sampling plan, a visual inspection was performed on thirteen samples.The packets were opened.The swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand.No anomalies were observed during the evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Two unused open samples were received for analysis.Product code ep8707h.In order to evaluate the conditions of the returned samples, the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE SUTURE 30"(75CM) 6-0 BLU
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22125881
Report Number2210968-2025-06113
Device Sequence Number7274042
Product Code GAW
UDI-Device Identifier10705031231306
UDI-Public(01)10705031231306(17)290531(10)101HA1
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEP8707H
Device Lot Number101HA1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2025
Supplement Date Received by Manufacturer06/09/2025
Initial Report FDA Received Date06/02/2025
Supplement Report FDA Received Date07/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-