|
|
| Catalog Number |
412.116S |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: d4: udi: as the catalog/model number was not provided, the (01) gtin is not available.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, the patient underwent orif surgery for proximal humerus fracture with plate and screws.In the surgery, seven screws were inserted proximally and three distally.The surgery was completed successfully without any surgical delay.After surgery, on (b)(6) 2025, it was confirmed that at least five proximal screws had been dislocated.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10 (concomitant).H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 (lot, primary udi #).Corrected: d1, d4 catalog.D4: udi: as the lot number for the device involved in the event was for dhr review, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 expiration date, h4.Corrected: d4 primary udi number, g1.H3, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there were no damage or defect with the lockscr ø3.5 self-tap l38 tan.An interactional test was performed with the plate returned in this comlaint and the screw was able to screwed as expected on the locking hole.In addition, the allegation of migration/backout or screws dislocated was not confirmed as post-operative images or x-ray to assess the condition were not provided.A dimensional inspection for the lockscr ø3.5 self-tap l38 tan was not performed as it is not applicable to the complaint condition.The overall complaint was not confirmed as the lockscr ø3.5 self-tap l38 tan.Was found to have no damage or defects.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There was no indication that a design or manufacturing issue contributed to the complaint.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device.Product code: 412.116ts-15, lot number: 31704p8.It was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 13 sep 2024, manufacturing site: jabil grenchen, expiry date: 01 sep 2029.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|