| Model Number |
VTICM5_13.2 (-5.50/1.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Visual Disturbances (2140); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a vault value of 1100um, elevated intraocular pressure, significant reduction of iridocorneal angles, pigment dispersion, glare/haloes.In a separate visit the lens was exchanged for a shorter length and the problem was resolved.The cause was reported as device.
|
| |
|
Additional Manufacturer Narrative
|
|
H6-clinical code: 4581-significant reduction of irido-corneal angles, pigment dispersion h6-investigation type 4110: lens work order search- no similar complaint event(s) within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h6- method code 3331: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work orders(s), including the suspected device, have been manufactured within established parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|
| |
|