| Model Number |
VICM5 13.2 (-9.50) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced an excessive vault and significant reduction of iridocorneal angles.Reportedly, "the vault was over 1,500 microns with fairly narrow angles".In a separate surgery, the lens was exchanged with the same model, shorter length and the problem was resolved.The cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H6- health effect- clinical code: 4581- significant reduction of iridocorneal angles.H6- investigation type code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.H11- manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim#(b)(4).
|
| |
|