• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, DISTAL MICROBORE TUBING, S; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, DISTAL MICROBORE TUBING, S; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 15247-28
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received for evaluation.Investigation is pending.D4: only di provided as lot number is unknown.
 
Event or Problem Description
The event occurred on an unspecified date and involved a primary plum set, clave secondary port, clave y-site, distal microbore tubing, secure lock, 104 inch where it was reported the iv tubing was leaking at vent port.The iv tubing was in use on a patient when leaking was noticed and pulled from use.There was no harm as a result of the event.
 
Additional Manufacturer Narrative
Received two used partial list #15247-28, primary plum set, clave secondary port, polyethylene-lined light resistant tubing, clave y-site, secure lock, 103 inch.As received, the filter vent was wetted out shows signs of leaking with prior infusate.The set was leak tested per specifications and could not replicate leakage from the filter vent on the returned sample.The complaint of a leak out of the vent located on the spike of the tubing above the drip chamber can be confirmed due to the presence of fluid residuals at received filter.The probable cause of leakage is due to a temporary or complete loss of hydrophobic properties of the filter vent material due to an infusate interaction during use.A dhr lot # review could not be conducted because no lot number was identified.Corrected information can be found in: -e3 - initial reporter occupation -e4 - initial report sent to fda -h6 - component code.
 
Additional Manufacturer Narrative
Corrected information: the previous supplemental should have been marked as both correction and device evaluation in h2 - if follow-up, what type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, DISTAL MICROBORE TUBING, S
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key22126220
Report Number9615050-2025-00272
Device Sequence Number15276916
Product Code FPA
UDI-Device Identifier00887709102057
UDI-Public00887709102057
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K141789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number15247-28
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2025
Supplement Date Received by Manufacturer06/23/2025
06/23/2025
Initial Report FDA Received Date06/02/2025
Supplement Report FDA Received Date07/18/2025
07/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-