| Model Number |
104.3605 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
05/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4) initial report.Additional information, including post primary and pre revision x-rays, operative notes (primary and revision), patient activity level, weight and medical history, whether the patient followed correct post-op protocol, whether the patient experienced any slips / falls or other trauma post primary, what the patient was doing at the time of the dislocation and an update on the patient following the revision, has been requested in order to progress with the investigation of this event, and if received, will be provided in a supplemental report upon completion of the investigation.The appropriate device details have been provided and the relevant device manufacturing records will be identified and reviewed.Please note: this report is filed with the fda due to an adverse event experienced with a device that is similar to those placed on the market in the usa, however, this event occurred outside of the usa.The submission of this report does not constitute an admission that the device, reporting entity, entity's representative or distributor caused or contributed to this event.
|
| |
|
Event or Problem Description
|
|
Trinity revision of the ecima liner and biolox delta ceramic head after approximately 2 months due to dislocation.
|
| |
|
Event or Problem Description
|
|
Trinity revision of the ecima liner and biolox delta ceramic head after approximately 2 months due to dislocation.
|
| |
|
Additional Manufacturer Narrative
|
|
Per - (b)(4) final report.Additional information, including post primary and pre revision x-rays, operative notes (primary and revision), patient activity level, weight and medical history, whether the patient followed correct post-op protocol, whether the patient experienced any slips / falls or other trauma post primary, what the patient was doing at the time of the dislocation and an update on the patient following the revision, was requested in order to progress with the investigation of this event, however, not all information could be provided and thus the scope of the investigation was limited.It was stated that the patient had a complicated medical history with comorbidities and it was unknown if they strictly followed post-op protocol, however, they did not report to have had any falls or other tramua post primary surgery.The appropriate device details were provided and the relevant device manufacturing records have been identified and reviewed.All parts associated with these records conformed to material and dimensional specification at the time of manufacture.Based on the available information, no further investigation can be conducted and the root cause of the reported dislocation could not be determined and thus this case is now considered closed.However, should any additional information be provided then this case may be re-opened for further investigation.Please note: this report is filed with the fda due to an adverse event experienced with a device that is similar to those placed on the market in the usa, however, this event occurred outside of the usa.The submission of this report does not constitute an admission that the device, reporting entity, entity's representative or distributor caused or contributed to this event.
|
| |
|
Search Alerts/Recalls
|
|