| Catalog Number |
IAB-05830-LWS |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "persistent possible helium loss 2 alarms.They are coming every 2-3 min".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) returned for investigation was a 40cc 8.0fr fiberoptix ultra intra-aortic balloon catheter (iabc) without the original packaging.The sample was returned in the ups shipping box and was in a sealed biohazard bag.Upon return, the teflon sheath was noted on the iabc; liquid blood was noted within the sheath sidearm.The one-way valve was tethered to the short driveline tubing.The iabc bladder was fully unwrapped.Dried blood was noted on the exterior surfaces of the returned iabc.No blood was noted within the helium pathway.The fos connector and cal key were examined.The fos gray connector was properly seated in the blue clamshell housing and both retaining tabs were intact.The center post of the fos was centered.The blue clamshell housing was examined, and no abnormalities were noted.The cal key was intact; no damage or abnormalities were noted to the cal key.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.The cal key and fos were connected to the iabp.The cal key was recognized.The fos was connected and the iabp zeroed the iabc.The pump status displayed "ok" indicating the fiber is fully intact.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory syringe.No abnormalities or debris were noted.The iabc was leak tested in accordance with testing methods from manufacturing procedure.An external leak was immediately detected from the bifurcate around the fos adhesive.Upon microscopic inspection, an external leak occurred from the fos adhesive bond at the bifurcate consistent with showing a void.No other leaks were detected.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.No resistance was noted; the guidewire was able to advance through the central lumen.No blood or debris was noted.The reported complaint for "persistent possible helium loss 2 alarms" is confirmed.During the investigation, an external helium leak was confirmed from the iabc bifurcate fos adhesive, which caused the reported complaint.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the leak at the bifurcate fos adhesive.The probable root cause of the bifurcate fos adhesive leak is manufacturing related.A corrective and preventive action (capa) has been initiated to address the issue.
|
| |
|
Event or Problem Description
|
|
It was reported "persistent possible helium loss 2 alarms.They are coming every 2-3 min".Additional information states that the iab catheter was removed and replaced.No patient injury or consequence reported.The patient's current condition is reported as "fine".
|
| |
|