| Model Number |
8884720247E |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that a hole in the balloon of the kangaroo replacement probe was observed.There was no reported harm.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) for lot number 2310012664 was reviewed and no anomalies related to the reported issue were observed.There were no samples returned for evaluation; therefore, the affected device could not be evaluated to confirm the reported issue.As such, a root cause could not be determined at this time.However, a corrective and preventative action has been opened to address the reported issue.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|