| Model Number |
DIU150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Eye Injury (1845)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section a2, a3, a3b, a4 and a5: unknown/ asked but not available.Section h3: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the preloaded intraocular lens (iol) was explanted from patient's left eye due to other disorders of the eye.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d-9: device available for evaluation: yes section d-9: device date returned to manufacturer: june 10, 2025 section h-3: device evaluated by manufacturer: yes device evaluation: visual inspection under magnification was performed on the suspect product and found the lens was cut in half with both halves stuck together and coated in viscoelastic residue.The lens was cleaned and inspected, revealing damage to the edge of the lens.No further issues were observed.No further evaluation was performed.No issues that could cause or contribute to the complaint issues reported were identified during product evaluation.The other issues observed during the product evaluation could not be confirmed to be related to the manufacturing or design process.Based on the complaint investigation results, the product was released within specifications.No product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|