• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 436013C
Medical Device Problem Codes Mechanical Problem (1384); Material Deformation (2976); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having posterior t12 c orpectomy.It was reported that the device stuck in mid expansion, was not expanding and collapsing as it should and stripping while attempts were made to expand and collapse the cage.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis of part# 436013c, lot# ca23a184 - visual and optical examination identified that it appears that one of the teeth of the helical gear has been chipped off.The metal deformation gives indication that the failure was the result of overload and misalignment.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM
Common Device Name
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key22129541
Report Number1030489-2025-02125
Device Sequence Number8371456
Product Code MQP
UDI-Device Identifier00643169663138
UDI-Public(01)00643169663138(11)231204(10)CA23A184
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K173125
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number436013C
Device Catalogue Number436013C
Device Lot NumberCA23A184
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2025
Supplement Date Received by Manufacturer10/24/2025
Initial Report FDA Received Date06/03/2025
Supplement Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age44 YR
Patient SexFemale
-
-