• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC Back to Search Results
Catalog Number 706.52
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that during surgery, an ophthalmic scissors was found to be blunt and could not be used.The procedure was completed after replacing the new one.The patient impact has not been reported.
 
Additional Manufacturer Narrative
Additional information provided in h.6., and h.11.The concerned instrument was not returned to the investigation site for evaluation.Therefore, no further assessment is possible and the investigation is concluded without identifying the root cause.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A non-conformance based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint.A sample was not received at the investigation site and not enough information was provided for further investigation.Therefore, the root cause for the customer complaint issue cannot be determined conclusively.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP SCISSORS
Common Device Name
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22131619
Report Number3003398873-2025-00307
Device Sequence Number8408398
Product Code HNF
UDI-Device Identifier07612717066885
UDI-Public(01)07612717066885(17)261130(11)231207(10)16DC7W
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number706.52
Device Lot Number16DC7W
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/27/2025
Supplement Date Received by Manufacturer06/26/2025
Initial Report FDA Received Date06/03/2025
Supplement Report FDA Received Date07/21/2025
Date Device Manufactured12/07/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-