MAQUET SAS MAQUET POWERLED II; LAMP, SURGICAL Back to Search Results
Model Number ARDPWT229064A
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
Getinge became aware of an issue with one of surgical lights- powerled ii.It was stated the paint was peeling from the headlight's handle.There was no injury reported, however we decided to report the issue in abundance of caution as any parts or particles falling off into sterile field or during procedure may cause contamination.
 
Additional Manufacturer Narrative
Event site name: (b)(6).Getinge became aware of an issue with one of surgical lights- powerled ii.It was stated the paint was peeling from the headlight's handle.There was no injury reported, however we decided to report the issue in abundance of caution as any parts or particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.A root cause analysis for paint peeling problem was performed by subject matter experts and concluded that all maquet sas products comply with: current version of iec 60601-1 general requirements for basic safety and essential performance.Current version of iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.Paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.Disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.To prevent any similar incident, it is recommended to avoid collisions between devices.Visual inspections during the cleaning allow to detect the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor paint chip can be repaired with touch up paint, nevertheless, the parts impacted by serious damage must be replaced.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET POWERLED II
Common Device Name
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key22131816
Report Number9710055-2025-0000147
Device Sequence Number7487036
Product Code FTD
UDI-Device Identifier03700712411411
UDI-Public(01)03700712411411
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARDPWT229064A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2025
Initial Report FDA Received Date06/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/13/2023
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-