| Model Number |
CL-932 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during suturing, preparing the suture after patient incision, the two needle detached from the suture.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant medical products: cl-932 - cl-932 polysorb* 1 vio 75cm hgs21 x36 - lot# a5a1644y cl-932 - cl-932 polysorb* 1 vio 75cm hgs21 x36 - lot # a4j2006y medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d3(mfr name and street 1), d4(udi), g3, h3, h6 correction: e3 was removed, e4 h3 evaluation summary: medtronic conducted an investigation based upon all information received.Fifty seven representative samples were available for eval uation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the needle detached.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|