MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 2700
Medical Device Problem Codes Insufficient Device Problem Information (3190); Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Dyspnea (1816); Insufficient Information (4580)
Date of Event 04/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned through implant patient registry that a 23mm 2700 aortic valve was explanted after an implant duration of 8 years, 7 months due to unknown reason.The explanted device was replaced with a 27mm 11500a aortic valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
 
Event or Problem Description
It was learned through implant patient registry that a 23mm 2700 aortic valve was explanted after an implant duration of 8 years, 7 months due to regurgitation.The patient presented with dyspnea.The explanted device was replaced with a 27mm 11500a aortic valve.The patient expired on pod # 12.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: b4, b5, b7, g3, g6, h2, h6 (health effect - clinical code, device code).
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6 (type of investigation, investigation findings, investigation conclusions) regurgitation identified post-procedurally has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Regurgitation occurring post-procedurally is most commonly related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing non-conformance.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including hyperlipidemia (hld) and chronic kidney disease (ckd).The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.H11: corrected data: h6 (component code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES, LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key22132918
Report Number2015691-2025-04375
Device Sequence Number8371574
Product Code DYE
UDI-Device Identifier00690103021608
UDI-Public(01)00690103021608(17)200411
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/11/2020
Device Model Number2700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer06/04/2025
06/28/2025
Initial Report FDA Received Date06/03/2025
Supplement Report FDA Received Date06/24/2025
07/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/12/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age78 YR
Patient SexMale
-
-