| Model Number |
ANGEL PRP SYSTEM CENTRIFUGE US (NEW) |
| Medical Device Problem Codes |
Use of Device Problem (1670); Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, it was reported by a sales representative via (b)(4) that an abs-10060 angel prp system centrifuge displayed different error codes.This was discovered during the case with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 complaint allegation is not confirmed.One unpackaged abs-10060 serial number (b)(6).Was returned for investigation.Functional testing results: 60ml of bovine blood with acd-a was processed on the device with standard settings at seven percent hematocrit.The device reported outputs of 35 ml platelet-poor plasma (ppp), 22 ml rbc, and 4 ml prp.The prp volume within the syringe was recorded as 3 ml.Aliquots of the wb and prp were analyzed for cbcs with the heska hematology analyzer (table 1).Please note that the prp had expected cellular foldx concentrations.The complaint describes that the system in question, sn (b)(6), displayed multiple error codes.This error could not be replicated.Visual evaluation noted no issues with the device.Manufacturing date: 21-nov-2024 no problem found.See attached testing investigation memo.Refer to investigation photos.The complaint allegation is not confirmed.
|
| |
|