Model Number PM2160
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Death
Event or Problem Description
It was reported that the patient has deceased.There is no known allegation from a health care professional that suggests that the death was device related.The cause of death was unknown.No additional information was reported.
 
Additional Manufacturer Narrative
Further information was requested but not received.
 
Additional Manufacturer Narrative
Device returned for analysis due to patient death.Analysis performed, including electrical, mechanical and environmental tests indicated that the device exhibited normal device characteristics with no anomalies.Longevity assessment was performed and device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY DR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key22134579
Report Number2017865-2025-75329
Device Sequence Number7445698
Product Code NVZ
UDI-Device Identifier05414734507059
UDI-Public(01)05414734507059(10)A000007330(17)161231
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Model NumberPM2160
Device Lot NumberA000007330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2025
Supplement Date Received by Manufacturer06/23/2025
Initial Report FDA Received Date06/03/2025
Supplement Report FDA Received Date07/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/04/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexMale
-
-