Catalog Number MWM052
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/07/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional information become available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the impactor tip was found broken in two pieces.
 
Additional Manufacturer Narrative
Correction: d5 operator of device.The reported event was confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: visual inspection: the visual inspection of the returned device reveals that the impactor tip has fractured into two separate pieces.Several external impact marks are evident on the tip's surface.The observed breakage pattern, originating from the specimen¿s edge and characterized by v-shaped features, suggests catastrophic failure likely due to rough handling or excessive application of mechanical force.A review of the device history for the reported lot did not indicate any abnormalities.This issue has been addressed by a corrective action.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.Most likely, the event was caused by application of excessive mechanical force.If more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that the impactor tip was found broken in two pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM HUMERAL IMPACTOR TIP SIZE 3/4
Common Device Name
IMPACTOR
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key22134859
Report Number0001649390-2025-00377
Device Sequence Number12864080
Product Code HWA
UDI-Device Identifier00846832082803
UDI-Public(01)00846832082803(11)231204(10)CV0123323
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMWM052
Device Lot NumberCV0123323
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/08/2025
Supplement Date Received by Manufacturer07/11/2025
Initial Report FDA Received Date06/03/2025
Supplement Report FDA Received Date08/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-