| Catalog Number |
MWM052 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional information become available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the impactor tip was found broken in two pieces.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d5 operator of device.The reported event was confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: visual inspection: the visual inspection of the returned device reveals that the impactor tip has fractured into two separate pieces.Several external impact marks are evident on the tip's surface.The observed breakage pattern, originating from the specimen¿s edge and characterized by v-shaped features, suggests catastrophic failure likely due to rough handling or excessive application of mechanical force.A review of the device history for the reported lot did not indicate any abnormalities.This issue has been addressed by a corrective action.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.Most likely, the event was caused by application of excessive mechanical force.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the impactor tip was found broken in two pieces.
|
| |
|
Search Alerts/Recalls
|