Model Number E2516H-GNSB
Medical Device Problem Codes Fire (1245); Arcing (2583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an intraoperative procedure, the electrode tip of the device was visibly melted, as also seen in the photo received.It was also reported that the device sparked.The settings on the ft10 unit at the time of the incident were blend and fulgurate power level of 30 watts.Additionally, testing was conducted with the hospital's biomed department, two pencils and two electrodes created sparks when the blend mode on the ft10 was set at 100.It was further reported that a flame with the size of a birthday candle was produced and was extinguished by releasing off the coagulation button.Biomed tested the generator and it was within specification and without issues.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
D10 concomitant products: vlft10gen, vlft10gen ft series energy platformx1 (serial #:(b)(6), e1455nsb, e1455nsb b the edge ent/ima electr x50 (lot#:231350286r), e2516h-gnsb, e2516h-gnsb e2516h-gnsb disposable hand (lot#:222830308r), e1455nsb, e1455nsb b the edge ent/ima electr x50 (lot#:231350286r), e1455nsb, e1455nsb b the edge ent/ima electr x50 (lot#:231350286r) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the electrode tip of the device was visibly melted and the device sparked.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22135043
Report Number1717344-2025-00715
Device Sequence Number8391580
Product Code GEI
UDI-Device Identifier10884524003016
UDI-Public10884524003016
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K813071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE2516H-GNSB
Device Catalogue NumberE2516H-GNSB
Device Lot Number222830308R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2025
Initial Report FDA Received Date06/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient SexUnknown
-
-