SMITH & NEPHEW, INC. BLADE,4.5 BONECUTTER,FR,180MM LG,DSP; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
7210979 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Material Twisted/Bent (2981); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the bonecutter blade had metal shavings/debris.All the shred material was removed from the patient with oscillation mode in the shaver.The procedure was completed using a back-up device.There was a delay of 1-5 minutes and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Factors that can contribute to the reported event include an application of unintended inappropriate or excessive force to the device.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for further assessment.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h2: corrected data on e1.
|
| |
|