| Model Number |
TJF-150 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.The event can be detected/prevented by following the instructions for use in: the detection method in tjf type 150 operation manual chapter 3 preparation and inspection.Ifu states the preventative measures in tjf type 150 reprocessing manual chapter 3 cleaning, disinfection and sterilization procedures.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed during the device evaluation that the duodenovideoscope exhibited foreign objects within the forceps elevator wire.There were no reports of patient involvement.
|
| |
|