• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Catalog Number 130790030
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Event or Problem Description
The reverse shoulder prosthesis was implanted in our hospital four years ago.Its function is excellent.A firm, elastic bulge has formed anteriorly in the area of the deltoideopectoral approach.Radiological examination on (b)(6) 2024, revealed a dislocation of the glenosphere relative to the metaglene.There is an indication for replacement of the glenosphere and, if necessary, also the metaglene in conjunction with an inlay replacement.The revision surgery took place on (b)(6) 2025.The procedure was completed with no issues noted.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: hcp is reporting to nca/bfarm: "the reverse shoulder prosthesis was implanted in our hospital four years ago.Its function is excellent.A firm, elastic bulge has formed anteriorly in the area of the deltoideopectoral approach.Radiological examination on (b)(6) 2024, revealed a dislocation of the glenosphere relative to the metaglene.There is an indication for replacement of the glenosphere and, if necessary, also the metaglene in conjunction with an inlay replacement.The planned surgery took place on (b)(6) 2025.The following findings were observed: (copy from the surgical report) ".This clears the path into the joint, and the articular membrane with severe metallosis is removed, both around the shaft and in the area of the glenosphere.This membrane has completely detached from the metaglene, but still articulates well with the shaft portion of the prosthesis.First, the glenosphere is removed, followed by the inlay.This provides even better access to the joint and allows the synovialectomy to continue.Only at the lower edge of the glenoid is one careful to avoid damaging the axillary nerve.Then, the screw residue of the glenosphere locking screw is exposed in the middle of the metaglene.Fortunately, it had a central hole that a conical male left-hand thread cutter engages.The metaglene is then cleanly freed of scar tissue all around, and a new 38 mm glenosphere is applied caudally with the eccentric tool.Fixed."intraoperatively, photo documentation done." the product was not returned to j&j medtech orthopaedics; however a photo was provided for review.(b)(4) the photograph attached was reviewed, however it does not represent the reported product complaint.Therefore, the investigation could not draw any conclusions about the reported product due to the insufficient evidence provided.A manufacturing record evaluation was performed for the finished device [130790030 / 5360310], and no non-conformances or manufacturing irregularities were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the photographs provided are not representative of the reported product and adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot:1) quantity manufactured: (b)(4) 2) date of manufacture: 1/29/2020.3) any anomalies or deviations identified in dhr: none.4) expiry date: 12/31/2024.5) ifu reference: (b)(4) device history review: a manufacturing record evaluation was performed for the finished device [130790030 / 5360310], and no non-conformances or manufacturing irregularities were identified.H11 additional narrative: added: d4 (exp.Date).Corrected: h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X30MM
Common Device Name
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22136062
Report Number1818910-2025-08966
Device Sequence Number8371401
Product Code HSD
UDI-Device Identifier10603295027843
UDI-Public(01)10603295027843(17)241231(10)5360310
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Catalogue Number130790030
Device Lot Number5360310
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer08/06/2025
Initial Report FDA Received Date06/03/2025
Supplement Report FDA Received Date08/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-