Catalog Number 730M0017
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during inspection of a field return from a hospital, a warehouse discovered a fractured tip on a vital t27 final driver.The hospital did not provide any patient or event information.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Additional Manufacturer Narrative
Corrections in d9 and h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip is fractured.Root cause: this event could possibly be attributed to excessive torsional forces applied to the tip during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that during inspection of a field return from a hospital, a warehouse discovered a fractured tip on a vital t27 final driver.The hospital did not provide any patient or event information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T27 FINAL DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key22136065
Report Number3012447612-2025-00241
Device Sequence Number18235421
Product Code HXX
UDI-Device Identifier00880304852761
UDI-Public(01)00880304852761(10)4517802
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number730M0017
Device Lot Number4517802
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/03/2025
Supplement Report FDA Received Date09/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/28/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-